3-1-4. Initial Setting
1. Switch settings When changing the following switches from a customer-set position, it is recommended to record the setting states in the table below. After adjustment is complete, be sure to return the switches to their customer-set positions.
3-2. D/A Y Level Adjustment
Equipment: Oscilloscope Adjustment Procedure Test point: GND: Adjustment point: Specification: TP10/DA-126 E3/DA-126 1RV1/DA-126 A = 1000 ± 10 mV